Immersion oil in microscopy plays a crucial role in achieving high-resolution images by reducing the amount of refraction and increasing the numerical aperture (NA) of the objective. The refractive index (RI) of the immersion medium, such as oil, determines the amount of refraction through the NA of the objective. A higher RI allows for a larger NA, which in turn collects more light and forms a brighter and sharper image.
Understanding Refractive Index and Numerical Aperture
The refractive index (RI) of a material is a dimensionless quantity that describes how the speed of light is reduced when it passes through that material. The RI is calculated as the ratio of the speed of light in a vacuum to the speed of light in the material.
The formula for the refractive index is:
n = c / v
Where:
– n
is the refractive index (dimensionless)
– c
is the speed of light in a vacuum (approximately 3 × 10^8 m/s)
– v
is the speed of light in the material
For instance, the RI of air is 1.0003, while the RI of oil and glass is 1.51. This means that light travels about 1.51-fold slower in oil and glass compared to air. When traveling between materials with different RIs, the light path will bend, a process known as refraction.
The numerical aperture (NA) of an objective is a measure of the range of angles over which the objective can accept or emit light. The NA is calculated using the formula:
NA = n × sin(θ)
Where:
– n
is the refractive index of the immersion medium
– θ
is the half-angle of the maximum cone of light that can enter or exit the objective
A higher NA indicates a larger area over which light can be captured, resulting in more light being collected and a brighter image.
Choosing the Appropriate Immersion Oil
The choice of immersion oil is crucial in achieving high-resolution images in microscopy. The RI of the immersion oil should match the RI of the glass through which the sample is visualized to minimize refraction and maximize the NA of the objective.
In a study comparing two different immersion oils, Type FF and Type LDF, it was found that using an immersion oil that matches the RI of the glass (Type LDF, RI = 1.51) resulted in improved optical image quality and greater dendritic spine densities compared to using an immersion oil that matches the RI of the tissue and mounting medium (Type FF, RI = 1.47).
In terms of numerical values, the RI of Type FF immersion oil is 1.47, while that of Type LDF immersion oil is 1.51. In this study, dendritic spine densities from sections imaged with Type LDF immersion oil were greater than those imaged with Type FF immersion oil in all three striatal subregions.
Viscosity and Specimen Movement
In addition to the RI, the viscosity of the immersion oil is also an important factor to consider. Water immersion objectives, for example, use water as the immersion medium instead of oil. Water has a lower viscosity than most immersion oils, which exerts less force (surface tension) on the cover glass during focusing. This minimizes specimen movement during the repeated refocusing required during optical sectioning, resulting in sharper and more meaningful three-dimensional reconstructions from the image stack.
The viscosity of the immersion medium can be measured in centipoise (cP) or millipascal-seconds (mPa·s), which are both units of dynamic viscosity. Typical values for immersion oils range from 200 to 500 cP, while water has a viscosity of 0.89 cP at 20°C.
Specialized Immersion Media
In addition to standard immersion oils, there are specialized immersion media available for specific applications in microscopy. These include:
- High Refractive Index Immersion Oils: These oils have RIs greater than 1.51, which can further increase the NA of the objective and improve image resolution. Examples include:
- Cargille Type DF immersion oil (RI = 1.53)
-
Cargille Type HI immersion oil (RI = 1.78)
-
Low Viscosity Immersion Fluids: These fluids have lower viscosities than standard immersion oils, which can help minimize specimen movement during optical sectioning. Examples include:
- Glycerol (RI = 1.47, viscosity = 1490 cP at 20°C)
-
Water (RI = 1.33, viscosity = 0.89 cP at 20°C)
-
Immersion Media for Specific Wavelengths: Some immersion media are designed to work optimally with specific wavelengths of light, such as ultraviolet (UV) or infrared (IR) light. These can be useful for specialized imaging techniques.
Practical Considerations
When using immersion oil in microscopy, it is important to consider the following practical aspects:
-
Compatibility with Microscope Components: Ensure that the immersion oil is compatible with the objective lens, cover glass, and other microscope components to avoid damage or degradation.
-
Refractive Index Matching: Carefully measure the RI of the cover glass, mounting medium, and sample to determine the appropriate immersion oil with a matching RI.
-
Minimizing Contamination: Avoid introducing air bubbles or other contaminants into the immersion oil, as this can degrade image quality. Clean the objective lens and cover glass thoroughly before applying the oil.
-
Proper Oil Application: Apply a small drop of immersion oil directly onto the center of the cover glass or objective lens, and gently lower the objective into the oil to avoid trapping air bubbles.
-
Cleaning and Maintenance: Clean the objective lens and cover glass after use, and store the immersion oil in a clean, airtight container to prevent evaporation and contamination.
By understanding the principles of refractive index, numerical aperture, and viscosity, and by carefully selecting and using the appropriate immersion medium, physics students can achieve high-resolution, high-quality images in their microscopy studies.
References:
- Sternberg, S. R. (1983). Biomedical image processing. Computer, 16(1), 22-34.
- Inoué, S., & Spring, K. R. (1997). Video microscopy: the fundamentals. Springer Science & Business Media.
- Pawley, J. B. (2006). Handbook of biological confocal microscopy. Springer Science & Business Media.
- Sheppard, C. J., & Török, P. (1997). Effects of specimen refractive index on confocal imaging. Journal of microscopy, 185(3), 366-374.
- Minsky, M. (1988). Memoir on inventing the confocal scanning microscope. Scanning, 10(4), 128-138.
The lambdageeks.com Core SME Team is a group of experienced subject matter experts from diverse scientific and technical fields including Physics, Chemistry, Technology,Electronics & Electrical Engineering, Automotive, Mechanical Engineering. Our team collaborates to create high-quality, well-researched articles on a wide range of science and technology topics for the lambdageeks.com website.